DULCOTEST ICT 5

Description product group
  • Measured variable: electrolytic conductivity. The inductive (non-contact) measuring principle permits applications in chemically contaminated types of water and in film-forming media
  • Complete with injection moulded PP sensor head, no apparent bonds, seals
  • Measurements at high conductivity values of up to 2,000 mS/cm are possible without interfering polarisation by means of the high measuring range dynamics of the inductive measuring principle
  • Simple installation in PVC pipework by bonding the DN 40 adhesive connector supplied into a standard T-piece and screwing in the sensor using the union nut supplied.
  • A DN 40 welded connector is optionally available for use in PP pipework

You can find more product details and downloads on our Corporate Website.

Properties / Your benefits

Measuring range

0.2…2,000 mS/cm

Cell constant

6.25 cm-1

Measuring accuracy

±2% based on the measured value ±30 μS/cm

Temperature sensor

Pt 1000, wetted material Stainless steel 1.4301

Medium temperature

-10...80 °C -10...60 °C for installation in PVC pipes, -10...80 °C for installation in PP pipes

Max. pressure

10.0 bar at 20 °C, 6.0 bar at 60 °C, 0.0 bar at 80 °C

Min. pressure

-0.1 bar (-10 … 80 °C)

Sensor material

PP

Seals

EPDM

Electrical Connection

10 m fixed cable, 7x 0.35 mm2 via a terminal

Enclosure rating

IP 65

Typical applications

Contaminated waste water, blowdown control in cooling towers, control of electroplating and rinsing baths, cleaning in Place (CIP), product monitoring, sea water, brine swimming pools.

Resistance to

Ingredients in the water of the target application, taking into account compatibility to PP/EPDM, deposit-forming media

Process integration

With union nut, PVC, 1 1/2 inch female thread, including DN 40 bonded nozzle with 1 1/2 inch male thread for fitting in DN 40 PVC standard pipes (included in the scope of delivery). The corresponding set-in nozzle for fitting in PP standard pipes is available as an accessory

Controllers

Compact controller DCCa

Measuring principle, technology

Inductive, 2 coils. Integrated temperature measurement

Delivery details
Delivery time: See per variant
Variants
Order Number Price Delivery time Measuring range
1095248
3-4 weeks 0.2…2,000 mS/cm
Order Number: 1095248
Cell constant :
6.25 cm-1
Degree of protection :
IP 65
Electrical connection :
10 m fixed cable, 7x 0.35 mm2 via a terminal
Max. pressure :
10.0 bar at 20 °C, 6.0 bar at 60 °C, 0.0 bar at 80 °C
Measured variable :
electrolytic conductivity. The inductive (non-contact) measuring principle permits applications in chemically contaminated types of water and in film-forming media
Measuring range :
0.2…2,000 mS/cm

 


Do you need a quote?

After logging in, place the desired items in the shopping cart and click on "Request a quote" or send us a request via our contact form.

The product you are looking for is not available in the WebShop?

We currently offer a limited product portfolio in our WebShop. If you cannot find the product you are looking for, send us a request or call us at +44 1530 560555.

Send Request